• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number N160
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439); Under-Sensing (1661)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The local area field representative was contacted for additional information regarding the device serial number and patient outcome.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited farfield signal oversensing and atrial undersensing with competitive pacing while in a mode switch.Technical services (ts) discussed troubleshooting options and programming considerations.Multiple attempts have been made to obtain additional information regarding the model/serial number and patient information without success.At this time no further information is available.If additional information becomes available this report would be updated at that time.Evidence suggests this crt-d remains in service and no adverse patient effects were reported.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited farfield signal oversensing and atrial undersensing with competitive pacing while in a mode switch.Technical services (ts) discussed troubleshooting options and programming considerations.Multiple attempts have been made to obtain additional information regarding the model/serial number and patient information without success.At this time no further information is available.If additional information becomes available this report would be updated at that time.Evidence suggests this crt-d remains in service and no adverse patient effects were reported.Additional information received reported that the observations associated with this cardiac resynchronization therapy defibrillator (crt-d) system were due to atrial flutter response (afr) having been programmed on.Furthermore, this crt-d was explanted and replaced due to normal battery depletion (nbd) later that day.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.Correction to d4: serial number was updated to (b)(6).Correction to d6a/d6b: implant date and explant date provided via medical records upon locating the serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18994203
Report Number2124215-2024-18889
Device Sequence Number14202398
Product Code LWP
UDI-Device Identifier00802526480997
UDI-Public00802526480997
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/11/2016
Device Model NumberN160
Device Catalogue NumberN160
Device Lot Number969259
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer04/17/2024
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date04/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-