| Model Number |
N160 |
| Medical Device Problem Codes |
Over-Sensing (1438); Pacing Problem (1439); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information regarding the device serial number and patient outcome.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited farfield signal oversensing and atrial undersensing with competitive pacing while in a mode switch.Technical services (ts) discussed troubleshooting options and programming considerations.Multiple attempts have been made to obtain additional information regarding the model/serial number and patient information without success.At this time no further information is available.If additional information becomes available this report would be updated at that time.Evidence suggests this crt-d remains in service and no adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited farfield signal oversensing and atrial undersensing with competitive pacing while in a mode switch.Technical services (ts) discussed troubleshooting options and programming considerations.Multiple attempts have been made to obtain additional information regarding the model/serial number and patient information without success.At this time no further information is available.If additional information becomes available this report would be updated at that time.Evidence suggests this crt-d remains in service and no adverse patient effects were reported.Additional information received reported that the observations associated with this cardiac resynchronization therapy defibrillator (crt-d) system were due to atrial flutter response (afr) having been programmed on.Furthermore, this crt-d was explanted and replaced due to normal battery depletion (nbd) later that day.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.Correction to d4: serial number was updated to (b)(6).Correction to d6a/d6b: implant date and explant date provided via medical records upon locating the serial number.
|
| |
|