| Model Number |
250 (+ 20 D) |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Event occurred in malaysia damaged haptic after implantation problem code: a0414, material split cut or torn trailing; middle; damaged haptic increased incision size; product replaced with another lens immediately during surgery patient health not impacted iol was explanted on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to fda for an event that occurred outside of the usa.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu).250: pkg-23-021 00 en3.Ms3.150151250251.20220501(t)_15000f,15100f,25000f,25100f.Regarding section h6 - manufacturer's codes for: type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|
Event or Problem Description
|
|
Event occurred in malaysia, damaged haptic after implantation, problem code: a0414, material split cut or torn, trailing; middle; damaged haptic, increased incision size; product replaced with another lens immediately during surgery patient health not impacted, iol was explanted on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report #1 emdr is being submitted to fda for a reportable event that occurred outside the usa.The report includes corrected and additional information not available/included in the initial report.Corrected information: h3 - corrected to yes additional information: g6 - type of report - noted as follow-up #1 h2 - type of follow-up - noted for additional information h6 - added codes for manufacturer's investigation: type; findings; and conclusion.The product was returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.Appearance check result was consistent to reported information.No abnormalities were found in production and inspection records of the product.(serial no.: (b)(6) ; model: "250").We also confirmed there were no abnormalities on the "loop pull strength test" of the material lot.("sovj-g115-29").The dye test result showed the injector tip was properly coated.The lens release test result showed that the re-installed iol could be released out of the returned cartridge/tip without any problems.From our investigation, we confirmed the reported event.The exact root cause was not determined.However, based on the available information and our investigation, we believe this event was not caused by our product quality.Capa-23-0017 has been initiated for damaged haptics complaints.
|
| |
|