• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXALT CONTROLLER; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXALT CONTROLLER; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00542430
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report number 3005099803-2024-01335 for the exalt model d controller, and 3005099803-2024-01375 for the exalt model d scope.It was reported to boston scientific corporation that an exalt model d controller was used on an endoscopic retrograde cholangiopancreatography (ercp) performed on an unknown date.It was reported that the exalt model d controller shut down mid procedure on multiple occasions.The controller was then powered on again to complete the procedure.There were no reported patient complications as a result of this event.No further information has been obtained despite good faith efforts.
 
Additional Manufacturer Narrative
Block b3: the exact date of the event is unknown.The provided event date, 02/01/2024, was chosen as a best estimate based on the date that the manufacturer became aware of the event, 02/28/2024.Block h6 (device codes): problem code a06 captures the reportable event of loss of visualization inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXALT CONTROLLER
Common Device Name
DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
ENERCON TECHNOLOGIES
25 northbrook dr
665
gray ME 04039
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18994589
Report Number3005099803-2024-01335
Device Sequence Number20734246
Product Code FDT
UDI-Device Identifier08714729990901
UDI-Public08714729990901
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K193202
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00542430
Device Catalogue Number4243
Device Lot Number2314600012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2024
Initial Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-