MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS EDGE PUMP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS EDGE PUMP; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66803126
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The device was returned for evaluation.Device event logs confirm that the pump had experienced a critical failure, which was unable to be resolved, logs also confirm the device failed the leak and over vacuum testing., during self test.Visual examination of the returned device confirmed that particulate contamination was evident on the air inlet port.A preliminary risk assessment ¿pra¿ was conducted.The pra was initiated following complaints for renasys edge failing self-test and noisy operation, the investigation recovered a number of the devices which were all inspected.The results concluded.Black particulate matter from the carbon filter component of the renasys edge canisters has been confirmed to be entering the npwt pump and causing damage.It is suspected the dust filter between the carbon filter and the connection to the pump inlet is failing/ not sufficiently bonded.The pra was escalated to health hazard evaluation ¿hhe¿ to assess the risk to the user.The hhe concluded.In the most likely scenario, the canister and pump operate normally.There is no hazardous situation and no harm, and the health risk index is a ¿ none/negligible risk region.In the worst-case scenario, the canister filter fails during home healthcare use and pump stops functioning.The pump alerts the user to failure or leak alarm.There may be an extended delay in therapy while a replacement pump is sourced.This could potentially lead to wound deterioration or skin graft failure.The health risk index is b ¿ low risk region.Based on the potential for cumulative effects of the canisters on the pump function, the hhe conclusion is to advance to field action board ¿fab¿.The field action board initiated a product recall on both renasys edge 300 & 800ml canisters with solidifiers.The root cause of the failure was determined to be that carbon particles from the canister odour filter had passed through the dust filter into the pump unit of the device, which affects its ability to provide suction and can result in excess noise.Non-conformance ¿nc¿ was created to capture the investigation and plan changes with a corrective action currently in place to manage the actions needed to reduce the probability of future occurrences of this nature.Smith and nephew will continue to monitor for adverse trends relating to the reported allegation.S+n is conservatively reporting this event in association with a class ii recall of the concomitant canisters pursuant to applicable regulations.
 
Event or Problem Description
It was reported that one (1) renasys edge pump failed the flow test:, the leak test and the over vacuum cut-off.As this was noticed in a non-therapeutic environment, there was not any patient involved.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Following a review of the available information, the manufacturer has determined that this event does not meet the criteria for mdr reporting.The device is intended to display an alarm/ alert if it detects an issue and may discontinue applying therapy until the issue is resolved.The loss of negative pressure wound therapy benefits would not be likely to cause or contribute to death or serious injury.Event may cause minor or temporary injury (e.G., delayed wound healing, maceration, etc.) requiring no or minor medical intervention.Event may require use of a backup device to continue therapy.This event did not lead to a death or serious injury, nor would it be likely to cause or contribute to a death or serious injury if it were to recur.This event is considered not reportable pursuant to applicable regulations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS EDGE PUMP
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18994805
Report Number8043484-2024-00030
Device Sequence Number14212516
Product Code OMP
UDI-Device Identifier05000223498634
UDI-Public05000223498634
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K223041
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number66803126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer01/14/2025
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date01/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES# 94248
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN RENASYS EDGE CANISTER
Patient SexUnknown
-
-