| Model Number |
8881850510 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
Customer reports: the product is leaking fluid out of the needle.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured between 11oct2022 and 12oct2022.Five opened but unused devices were returned for evaluation.Functional testing was completed per procedure; the reported issue was not observed on the returned product.Based on the available information, the reported condition could not be confirmed.A root cause was not determined.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.
|
| |
|