| Model Number |
IBAL UKA TIB CEMENTED TRAY, SIZE 1 LM/RL |
| Medical Device Problem Codes |
Positioning Failure (1158); Fitting Problem (2183)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 03/13/2024, it was reported by a sales representative via e-mail that ((2) ar-521-tba8 ibalance® uka tibial bearing implants were not locking into the ar-511-t1l ibalance® uka tibial tray.This occurred during a case.No additional information was provided, and additional information has been asked.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error.
|
| |
|