• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA LED®; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA LED®; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARDLCA209012A
Medical Device Problem Codes Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event site name: (b)(6) event site postal code: (b)(6).Additional information will be provided following the conclusion of the investigation.
 
Event or Problem Description
On 26th march, 2024 getinge became aware of an issue with one of surgical lights - lucea 50.It was stated the cap on the spring arm was missing.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination.
 
Event or Problem Description
On 26th march, 2024 getinge became aware of an issue with one of surgical lights - lucea 50.During maintenance it was found the cap on the spring arm was missing and the label was chipping.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
The correction of b5 describe event and problem, d1 brand name, d2a product code, previous d2b product code description, d4 catalog #, d4 unique identifier (udi) # and h4 manufacture date deems required.This is based on the internal evaluation.Previous b5 describe event and problem: on 26th march, 2024 getinge became aware of an issue with one of surgical lights - lucea 50.During maintenance it was found the cap on the spring arm was missing.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination.Corrected b5 describe event and problem: on 26th march, 2024 getinge became aware of an issue with one of surgical lights - lucea 50.During maintenance it was found the cap on the spring arm was missing and the label was chipping.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination.Previous d1 brand name: lucea 50/100.Corrected d1 brand name: lucea led®.Previous d2a product code: ftd.Corrected d2a product code: fsy.Previous d2b product code description.: lamp, surgical.Corrected d2b product code description.: light, surgical, ceiling mounted.Previous d4 catalog # ardlca209012a.Corrected d4 catalog # none.Previous d4 unique identifier (udi) #: n/a.Corrected d4 unique identifier (udi) #: (b)(4).Previous h4 manufacture date: 2019-07-10.Corrected h4 manufacture date: 2019-07-11.Getinge became aware of an issue with one of surgical lights - lucea 50.During maintenance it was found the cap on the spring arm was missing and the label was chipping.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination.It was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.The claimed device was not being used for patient treatment or diagnosis when the event took place.The most probable root cause of the break of this cap is repeated and violent shocks during the use of the device.Another probable root cause is that the cap has been forgotten or deteriorated after a re-adjustment of the spring arm during the maintenance of medical device.The yearly preventive maintenance program documented in the technical manuals mentions to check the presence and fixing of the plastic covers.The cap must be reinstalled during installation or after the maintenance procedure.We strongly advise to check similar devices in the hospital in order to check the presence of all spring arms caps.If a missing cap is noticed, a new one should be ordered as spare parts.Label peeling is probably due to collision or repeated excessive rubbing during cleaning of the device, the use of aggressive or unsuitable cleaning agents may be a contributing factor.The user manual mentions to check the lightheads for chipped paint, impact marks and any other damages during daily checks.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA LED®
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key18996306
Report Number9710055-2024-00233
Device Sequence Number2946408
Product Code FSY
UDI-Device Identifier03700712400637
UDI-Public(01)03700712400637
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K113679
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberARDLCA209012A
Device Catalogue NumberARDLCA209012A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer09/16/2024
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date09/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-