| Model Number |
11060A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Insufficient Information (4580)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10.Additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
Edwards received notification that a konect model 11060a23 was explanted the same day of implant due to dehiscence from the annulus of the patient.There was no allegation of device malfunction.Another konect model 11060a size 21 was successfully implanted in replacement.As reported, patient had significant blood loss, heart arrested after decannulation and the team decided not to resuscitate further and patient died.There was no allegation that the device was implicated.The surgeon was very experienced with konect.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Device dehiscence or suture torn out may occur early or late.When it occurs in the intraoperative period, it is typically a result of inadequate device implantation in combination with friable myocardial tissue.Valve dehiscence is not a malfunction related to a manufacturing deficiency of the device.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.
|
| |
|
Event or Problem Description
|
|
Edwards received notification that a konect model 11060a23 was explanted after an unknown duration due to dehiscence from the annulus of the patient.There was no allegation of device malfunction.Another konect model 11060a size 21 was successfully implanted in replacement.As reported, patient had significant blood loss, heart arrested after decannulation and the team decided not to resuscitate further and patient died.There was no allegation that the device was implicated.The surgeon was very experienced with konect.
|
| |
|
Event or Problem Description
|
|
Edwards received notification that a konect model 11060a23 was explanted at implant due to dehiscence from the annulus of the patient.It was successfully replaced before decannulation with another konect model 11060a, size 21.A concomitant coronary artery bypass was performed.As reported, the patient had significant blood loss, and heart arrested after decannulation.The team decided not to resuscitate further, and the patient died.There was no allegation that the device was implicated or malfunctioned.The surgeon was very experienced with konect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received through investigation from implant patient registry (ipr).
|
| |
|