| Model Number |
HINDFOOT NAIL CABLE TENSIONER |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via sems (b)(4) that an ar-9091-02 hindfoot nail cable tensioner screw is stripped.It was hard to tighten and loosen.Was damaged when putting it on the jig.As soon as it was placed on the jig it popped off.This was discovered during a procedure with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: complaint is not confirmed.No related problem found.One unpackaged ar-9091-02 was received for evaluation.Visual inspection found that one bolt head and the device's body were damaged.The device was tested with a good set, and no problem was found.The most likely cause(s) for the reported failure is user error, as no problem was found when the device was functionally tested and evaluated with a good set.
|
| |
|