MAUDE Adverse Event Report: ARTHREX, INC. HINDFOOT NAIL CABLE TENSIONER; INTRAMED FIXATN ROD AND ACCESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. HINDFOOT NAIL CABLE TENSIONER; INTRAMED FIXATN ROD AND ACCESS Back to Search Results
Model Number HINDFOOT NAIL CABLE TENSIONER
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2024, it was reported by a sales representative via sems (b)(4) that an ar-9091-02 hindfoot nail cable tensioner screw is stripped.It was hard to tighten and loosen.Was damaged when putting it on the jig.As soon as it was placed on the jig it popped off.This was discovered during a procedure with no patient harm.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Additional information: complaint is not confirmed.No related problem found.One unpackaged ar-9091-02 was received for evaluation.Visual inspection found that one bolt head and the device's body were damaged.The device was tested with a good set, and no problem was found.The most likely cause(s) for the reported failure is user error, as no problem was found when the device was functionally tested and evaluated with a good set.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HINDFOOT NAIL CABLE TENSIONER
Common Device Name
INTRAMED FIXATN ROD AND ACCESS
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18997293
Report Number1220246-2024-01710
Device Sequence Number11471368
Product Code HSB
UDI-Device Identifier00888867335301
UDI-Public00888867335301
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K221031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHINDFOOT NAIL CABLE TENSIONER
Device Catalogue NumberAR-9091-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/19/2024
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-