| Model Number |
UNIVERS VAULTLOCK GLENOID TRIAL, LARGE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 3/13/2024, it was reported by a sales representative via sems-06512361 that an ar-9236-03pp univers vaultlock glenoid trial central peg broke off in the patient.All broken fragments were removed.This was discovered during a procedure on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged ar-9236-03pp serial/batch number (b)(6).Was received for investigation.No functional test was performed, as the damage to the device will prevent it from working as intended.Visual evaluation noted that the central peg has broken off.Several other grooves and holes were noted on the body of the device.User error is the most likely cause(s) for the reported and observed failure.Hitting the device with another device or object, prying/leveraging, or excessive forces applied during use, could be likely causes as well.
|
| |
|