Model Number UNIVERS VAULTLOCK GLENOID TRIAL, LARGE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 3/13/2024, it was reported by a sales representative via sems-06512361 that an ar-9236-03pp univers vaultlock glenoid trial central peg broke off in the patient.All broken fragments were removed.This was discovered during a procedure on (b)(6) 2024.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged ar-9236-03pp serial/batch number (b)(6).Was received for investigation.No functional test was performed, as the damage to the device will prevent it from working as intended.Visual evaluation noted that the central peg has broken off.Several other grooves and holes were noted on the body of the device.User error is the most likely cause(s) for the reported and observed failure.Hitting the device with another device or object, prying/leveraging, or excessive forces applied during use, could be likely causes as well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS VAULTLOCK GLENOID TRIAL, LARGE
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18997779
Report Number1220246-2024-01705
Device Sequence Number19556798
Product Code KWS
UDI-Device Identifier00888867250697
UDI-Public00888867250697
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K161108
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUNIVERS VAULTLOCK GLENOID TRIAL, LARGE
Device Catalogue NumberAR-9236-03PP
Device Lot NumberS569GT000
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer07/22/2024
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date08/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/19/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-