BIOMET 3I OSSEOTITE CERTAIN 2 IMPLANT 4 X 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIFOSS410
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced implant fracture at tooth site #44.Therefore, the implant was removed.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A2: patient age and date of birth unknown / not provided.A4: patient weight unknown / not provided.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 3252.H6: investigation conclusions code was added: 4307.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation was performed, damage to the implant was identified.The implant was fractured at the collar.The fractured piece was returned.Also, a screw identified to be fractured inside the implant.Dhr review was completed for the subject lot number 2016070938.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2016070938 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was the clinician selects implant that is unable to resist long-term occlusal loading.Therefore, based on the available information, device malfunction did occur.The implant was fractured at the collar.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE CERTAIN 2 IMPLANT 4 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18997949
Report Number0001038806-2024-00589
Device Sequence Number2867580
Product Code DZE
UDI-Device Identifier00844868023142
UDI-Public(01)00844868023142(17)210722(10)2016070938
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K100724
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2021
Device Catalogue NumberXIFOSS410
Device Lot Number2016070938
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date08/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/21/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient EthnicityNon Hispanic
-
-