Model Number M00534290
Medical Device Problem Codes Break (1069); Use of Device Problem (1670); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Imdrf device code a0401 captures the reportable event of a guide catheter break.Imdrf impact code f2301 captures the additional device to remove the broken guide catheter.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix biliary stent with naviflex delivery system was to be used in the common bile duct to treat a stricture secondary to cholangiocarcinoma during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.The patient's anatomy was tight and was dilated prior to stent placement.During the procedure, the guide catheter broke from the push catheter and was left in the common bile duct.The broken guide catheter was removed from the patient using graspers, and the procedure was completed with another of the same device.There were no patient complications reported as a result of this event.Note: it was reported that the guidewire was inside the patient during the attempted deployment.However, the advanix biliary stent with naviflex rx delivery system instructions for use (ifu) states, "for preloaded system only: if the guidewire is not completely retracted into the endoscope, the stent can not be fully deployed." the physician did not follow the steps cited in the ifu.It was reported that the barb flap cover was not used to insert the device through the biopsy cap.However, the advanix biliary stent with naviflex rx delivery system instructions for use (ifu) states, "pull back on the pull wire cap to adjust the length of the guide catheter in front of the stent, and engage the naviflex rx delivery system locking mechanism.Slide the stent barb cover so it is on the distal end of the blue push catheter near the stent." the physician did not follow the steps cited in the ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18998197
Report Number3005099803-2024-01295
Device Sequence Number19550048
Product Code FGE
UDI-Device Identifier08714729787389
UDI-Public08714729787389
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K101314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2025
Device Model NumberM00534290
Device Catalogue Number3429
Device Lot Number0031978948
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-