| Model Number |
M00510870 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: imdrf device code a0406 captures the reportable investigation finding of side car rx pushback.H10: the returned trapezoid rx basket was analyzed, and a product analysis observed the side car rx was torn and pushed back.The reported event was confirmed.With all the available information, the problems of side car rx could have occurred due to excessive manipulation or force applied to the handle when trying to open the basket.Additionally, if a guidewire was used through the side car rx and got stuck, attempts to pull it out could cause the side car rx to tear or become pushed back.Therefore, the most probable root cause is adverse event related to procedure.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a trapezoid rx was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During procedure, the sheath of the trapezoid rx was broken.Another trapezoid rx was used to complete the procedure.There were no patient complications as a result of this event.This event has been deemed a reportable event based on the investigation finding of side car rx pushback.Please see block h10 for full investigation details.
|
| |
|