| Catalog Number |
81324 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron select sps g124, they allege that did not have enough water coming out and the handpiece got hot, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Notes: 04/17/24 repair tech: dts.Emv hp ;cable, conn/gun cable open.Intermittent or no operation due to an open condition in the handpiece cable.Replaced damaged/worn components and recalibrated unit to factory specs.Qa-tl.Unit came in missing bottom screws.Handpiece (b)(6) 2023, old hdpc (b)(6) 2023, new hdpc 10/2023.
|
| |
|