| Model Number |
PED2-375-10 |
| Medical Device Problem Codes |
Positioning Failure (1158); Retraction Problem (1536); Malposition of Device (2616); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id unk-nv-fg (lot: unknown); product type: ; implant date n/a; explant date n/a.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the pipeline encountered resistance in the catheter.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm of the left internal carotid artery (ica) between the pcom and ophthalmic with a max diameter of 2.5mm and a 2.5mm neck diameter.The landing zone was 3.6mm distally and 3.8mm proximally.It was noted the patient's vessel tortuosity was normal.It was reported that they started deployment in the m2 exposing braid and it opened up nicely like normal, then the doctor dragged the device down to the terminus ended up too proximal of the desired distal landing zone.Needing to adjust, the doctor recaptured the tip coil, then adjusted his position and they tried to un-sheath tip coil.The doctor said he had very hard resistance and the device would not re-deploy out of the phenom 27.He pulled out the phenom 27 system with the device still inside.The doctor started over with new phenom 27 and new device 3.75x14 which deployed like normal with no issues.It was noted that the pipeline was stuck in the catheter in the distal section during resheathing.The catheter was flushed continuously with heparanized saline.The physician released the load in the system in attempt to resolve the issue, however, the issue was not resolved.It was noted that the distal section of the catheter was damaged as the pipeline was lodged at the distal end of the phenom 27 and was noted it may be kinked.The pushwire was not damaged.The pushwire/capture coil was stuck during retraction in the distal end of the phenom.The pushwire was not rotated during removal.The pipeline was used for an indication that is approved (on-label).The reported devices and any accessory devices prepared as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.Ancillary devices include a sheath, navien 58 guide catheter, phenom 27 microcatheter, and synchro select guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: ¿ as found condition: the pipeline flex shield was returned stuck inside the phenom 27 catheter; within the outer carton; inside of a sealed bio-hazard bag and a shipping box.¿ damage location details: the distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.The distal and proximal ends of the braid were opened and frayed.No bends were found on the pushwire.¿ testing/analysis: the pipeline flex shield was pushed out from the catheter lumen with difficulty.¿ conclusion: based on the returned devices, the customer complaint was confirmed as the pipeline flex shield was returned stuck inside the catheter.The pipeline flex shield and catheter were damaged.From the damages seen on the catheter (accordioning), braid (fraying), and hypotube (stretching); it appears there was a high force used.It is possible these damages occurred when the customer attempted to advance the pipeline flex shield through the catheter against resistance.However, the cause of resistance could not be determined.Possible cause of resistance includes lack of continuous flush with heparinized saline during the procedure.Customer reported that vessel tortuosity was normal and the continuous flush with heparinized saline was used during procedure., ruling out vessel tortuosity and continuous flush as potential causes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that the device did not jump during deployment.The tip of the catheter was not moved during deployment.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|