• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SV CATHETER 3F FULL TRAY (SINGLE-LUMEN); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SV CATHETER 3F FULL TRAY (SINGLE-LUMEN); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.
 
Event or Problem Description
It was reported, ¿picc line had to be required as it was found to have a hole in the side wall inside the vessel.No patient impact.¿ no other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak is confirmed and was determined to be use related.The product returned for evaluation was a 3fr powerpicc sv with a needleless injection cap.The catheter shaft was split into two pieces between the 5cm and 6cm depth marking.Use residue was observed on the catheter.An attempt to flush the catheter with water using a 12ml syringe revealed the proximal portion of the catheter was patent to infusion and aspiration with no observed leaks.An attempt to flush the distal catheter segment revealed a leak between the 7cm and 8cm depth marking.Tactile evaluation of the leak site revealed some kinking and deformation in the vicinity of the split.Microscopic inspection of the leak site revealed a longitudinally aligned split.The surface of the split was granular.Microscopic inspection of the shared break surfaces revealed glossy, striated surfaces which were consistent with a sharp instrument cut.The granular surface of the split and kinking of the catheter shaft suggests the damage likely occurred due to fatigue failure from repetitive mechanical stress.This complaint will be recorded for future trending and monitoring purposes.H3 other text : evaluation findings are in section h11.
 
Event or Problem Description
It was reported, ¿picc line had to be required as it was found to have a hole in the side wall inside the vessel.No patient impact.¿ no other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SV CATHETER 3F FULL TRAY (SINGLE-LUMEN)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key19000784
Report Number3006260740-2024-01346
Device Sequence Number16912676
Product Code LJS
UDI-Device Identifier00801741027529
UDI-Public(01)00801741027529
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102159
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number3173108
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/14/2024
Initial Report FDA Received Date03/28/2024
Supplement Report FDA Received Date05/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-