| Model Number |
N/A |
| Medical Device Problem Codes |
Fracture (1260); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, an eight hole plate would not lock and cut the band after repeated attempts.Eventually, the band just snapped per the surgeon¿s explanation.There was no harm to the patient.The surgical team opened another 360 kit and finished the sternal closure with it.The case was delayed less than 10 minutes with no adverse effects.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.A visual inspection was conducted on the returned sl360 system.The packaging was returned opened.Visual inspection shows signs of attempted product use including marking and scratching on the product as well as dried fluid on the product.The implant locked as intended.Further inspection of the product band shows that the band has snapped.The complaint is confirmed.A determination cannot be made as to what caused the band to snap.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The reported event is confirmed, based on product return.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|