| Model Number |
CR2 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device delivered a monophasic shock instead of a biphasic shock.In this state, wrong defibrillation therapy may be delivered.There was no report of patient use associated to the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device delivered a monophasic shock instead of a biphasic shock.In this state, wrong defibrillation therapy may be delivered.There was no report of patient use associated to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker determined that a blown relay was the cause of the reported issue due to electrical overstress.Stryker archived the device, and the customer received a replacement device.
|
| |
|