WILSON-COOK MEDICAL INC CAPTURA BIOPSY FORCEPS WITHOUT SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT Back to Search Results
Catalog Number DBF-2.4-230-S
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
In preparation for a colonoscopy examination, the user selected a cook captura biopsy forceps without spike.It was reported that the nurse opened the outer packaging of the disposable biopsy forceps and inspected the biopsy forceps head.When the colonoscope was used, it was discovered the forceps head was bent and the forceps flap was stuck in open position.The procedure was completed with another same device.This occurred prior to patient contact; there was no impact to the patient.
 
Additional Manufacturer Narrative
Continued: investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Note: exercising the handle while the forceps is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura biopsy forceps without spike are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA BIOPSY FORCEPS WITHOUT SPIKE
Common Device Name
PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
WILSON COOK MEDICAL INC.
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
8002454707
MDR Report Key19002687
Report Number1037905-2024-00186
Device Sequence Number18268138
Product Code PTS
UDI-Device Identifier00827002315761
UDI-Public(01)00827002315761(17)260607(10)W4734655
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/07/2026
Device Catalogue NumberDBF-2.4-230-S
Device Lot NumberW4734655
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2024
Initial Report FDA Received Date03/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-