MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM C; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM C; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: 010000661.Item name g7 pps ltd acet shell 48c.Lot # 7599630.E1: establishment name is (b)(6) medical university.G2: foreign: china customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : device in process to return.
 
Event or Problem Description
It was reported that the liner would not assemble mating with the cup.After taking a series of measures such as cleaning up the soft tissue and checking whether the screws are flush, the liner still could not be locked properly, and it was replaced with the corresponding ceramic liner to complete the surgery.There was a 10 mins delay.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d4; g3; h2; h3; h4; h6.Visual examination of the returned liner product/provided pictures identified there is damage to the scallops.No other measurements were taken due to use and damage.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.No product return/pictures were provided of the shell used in the procedure.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 32MM C
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19002856
Report Number0001825034-2024-00866
Device Sequence Number15816912
Product Code PBI
UDI-Device Identifier00880304526334
UDI-Public(01)00880304526334(17)270930(10)7358346
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000847
Device Lot Number7358346
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer07/11/2024
Initial Report FDA Received Date03/29/2024
Supplement Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
-
-