| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 010000661.Item name g7 pps ltd acet shell 48c.Lot # 7599630.E1: establishment name is (b)(6) medical university.G2: foreign: china customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : device in process to return.
|
| |
|
Event or Problem Description
|
|
It was reported that the liner would not assemble mating with the cup.After taking a series of measures such as cleaning up the soft tissue and checking whether the screws are flush, the liner still could not be locked properly, and it was replaced with the corresponding ceramic liner to complete the surgery.There was a 10 mins delay.There is no additional information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d4; g3; h2; h3; h4; h6.Visual examination of the returned liner product/provided pictures identified there is damage to the scallops.No other measurements were taken due to use and damage.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.No product return/pictures were provided of the shell used in the procedure.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information at the time of this report.
|
| |
|