KAZ USA, INC., A HELEN OF TROY COMPANY BRAUN; FOREHEAD THERMOMETER Back to Search Results
Model Number BFH175US
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Fever (1858)
Date of Event 03/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Kaz usa, inc.Has requested that the product be returned to our company for testing, but it has not yet been received.
 
Event or Problem Description
A consumer reported that her thermometer had allegedly given multiple false negative readings on her infant.The device allegedly gave readings that were consistantly 4.6-4.8°f low, and a fever of 100.8°f was later confirmed by a doctor.There were no complications from this incident, and the infant is doing well now.Kaz usa, inc.Has requested that the product be returned to our company for testing.
 
Event or Problem Description
A consumer reported that her thermometer had allegedly given multiple false negative readings on her infant.The device allegedly gave readings that were consistantly 4.6-4.8°f low, and a fever of 100.8°f was later confirmed by a doctor.There were no complications from this incident, and the infant is doing well now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAUN
Common Device Name
FOREHEAD THERMOMETER
Manufacturer (Section D)
KAZ USA, INC., A HELEN OF TROY COMPANY
marlborough MA 01752
Manufacturer Contact
bethany lee espie
400 donald lynch boulevard
suite 300
marlborough, MA 01752
MDR Report Key19002876
Report Number1314800-2024-00008
Device Sequence Number9429438
Product Code FLL
UDI-Device Identifier00328785001679
UDI-Public00328785001679
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K181015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberBFH175US
Device Lot Number02121TAV
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer03/07/2024
Initial Report FDA Received Date03/29/2024
Supplement Report FDA Received Date05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age4 MO
Patient SexUnknown
-
-