| Catalog Number |
J324H |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Purulent Discharge (1812); Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The single complaint was reported with multiple events.There are no additional details regarding the additional patient events.Attempts are being made to obtain the information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure date and name of index surgical procedure? the diagnosis and indication for the index surgical procedure? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? please describe the patient manifestations of the reported infection/reaction (location, severity, appearance, systemic or local infection).Please provide the onset date/time of infection/reaction from the initial surgical procedure.Were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? were cultures performed? if so, please provide the results.How much and what type of drainage is present in this wound? please describe any medical intervention performed including medication name and results.Were any pre-op cleansing procedures changed recently? if yes, please describe.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? what symptoms did the patient experience following the index surgical procedure? onset date? other relevant patient history/concomitant medications? does the patient have a known allergic history to any medical devices, food and/or medication? was allergy testing performed? if so, please describe with results.What is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? are there any photos available?.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a throat surgery on an unknown date and suture was used.Post-op, the patient experienced a reaction to the suture in the form of infection (pus).Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem.Additional h3 investigation summary: the product was returned to ethicon for evaluation.The returned product determined that it was received, seventeen unopened packets that pertain to the product code j324h.As per the sampling plan, a visual inspection was performed on thirteen samples.The packets were examined for visual defects: appearance, color package, wrinkles in the seal area, continuous seals, seal margins, over-sealing, damage (torn, punctured) and none was observed.The packets were opened, and the swage and attachment area were noted to be as expected.During the visual inspection, the sutures were correctly placed on the winding former.Sutures were dispensed without problems and examined along the strand and no anomalies or damage on suture surface could be observed.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached on the cause of the reported event.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|