MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. 300MM VBT CORD; THE TETHER - VERTEBRAL BODY TETHERING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. 300MM VBT CORD; THE TETHER - VERTEBRAL BODY TETHERING SYSTEM Back to Search Results
Catalog Number 204H0300
Medical Device Problem Codes Break (1069); Unintended Movement (3026)
Health Effect - Clinical Code Spinal Column Injury (2081)
Date of Event 03/28/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported reported a revision surgery was performed to address curve progression and suspected tether cord breakage.A posterior screw construct was added during the revision without removing the tether construct.This is report one of five for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Additional Manufacturer Narrative
H6: additional method code: 4110.Corrections in g1.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos were provided, so an evaluation is unable to be performed.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference reports 3012447612-2024-00058 through 3012447612-2024-00062.
 
Event or Problem Description
It was reported reported a revision surgery was performed to address curve progression and suspected tether cord breakage.A posterior screw construct was added during the revision without removing the tether construct.This is report one of five for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
300MM VBT CORD
Common Device Name
THE TETHER - VERTEBRAL BODY TETHERING SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key19003520
Report Number3012447612-2024-00058
Device Sequence Number13121294
Product Code QHP
UDI-Device Identifier00880304865068
UDI-Public(01)00880304865068(17)230831(10)2966738
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
H190005
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/30/2023
Device Catalogue Number204H0300
Device Lot Number2966738
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/29/2024
Supplement Report FDA Received Date09/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age19 YR
Patient SexFemale
Patient Weight59 KG
Patient EthnicityNon Hispanic
-
-