• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN IMMUNOLOGICAL TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN IMMUNOLOGICAL TEST SYSTEM Back to Search Results
Catalog Number 07027273190
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter complained of a discrepant low result for 1 patient sample tested for elecsys ferritin (ferritin) on a cobas e 801 analytical unit.The initial result was 2.07 ug/l.The initial result was not believed and the sample was repeated.The repeat result was 278 ug/l.This result was believed to be correct.
 
Additional Manufacturer Narrative
The e801 analyzer serial number was (b)(6).
 
Additional Manufacturer Narrative
Calibration and qc were acceptable."abnormal aspiration" and "sample foam detection" alarms were observed on the alarm trace data; these can indicate sample quality issues.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FERRITIN
Common Device Name
FERRITIN IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19003604
Report Number1823260-2024-00954
Device Sequence Number9383867
Product Code JMG
UDI-Device Identifier04015630939688
UDI-Public04015630939688
Combination Product (Y/N)Y
Initial Reporter CountryNL
PMA/510(K) Number
K971833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue Number07027273190
Device Lot Number76245601
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer04/29/2024
Initial Report FDA Received Date03/29/2024
Supplement Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-