| Model Number |
N/A |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 medical devices: unknown persona femoral unknown persona tibial insert.G2 foreign source: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent a right knee revision approximately four months post-implantation due to tibial tray loosening.No additional patient consequences were reported.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.D10 medical devices: femur trabecular metal cruciate retaining (cr) standard porous catalog#: 42502806202 lot#: 64708880.Articular surface fixed bearing cruciate retaining (cr) right 14 mm height catalog#: 42522000514 lot#: 65815867.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Updated: b4, g3, g6, h2, h3, h6, and h11.Reported event was confirmed by review of medical records.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: moderate to severe medial and lateral compartment degenerative change with joint space narrowing; osteophyosis; no fracture; eventual loosening of the tibial component.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|