• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE COPE NEPHROURETEROSTOMY SET; FAD STENT, URETERAL Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stiffener of an ultrathane cope nephroureterostomy set was difficult to remove from the drainage catheter.Additionally, the string that forms the proximal pigtail became attached when pulled.At this time, no adverse effects or additional procedures for the patient were reported due to this occurrence.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1- customer (person): postal code = (b)(6).E3- occupation: operating room clinical purchasing secretary g4- pma/510(k) #: k171603 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation.It was reported that the stiffener of an ultrathane cope nephroureterostomy set was difficult to remove from the drainage catheter.Additionally, the string that forms the proximal pigtail became detached when pulled.No adverse effects or additional procedures for the patient were reported due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, quality control procedures, and specifications, as well as a visual inspection, dimensional verification, and functional test of the returned device, were conducted during the investigation.A total of two blue flexible stiffeners along with one ultrathane cope nephroureterostomy catheter were returned in a used and damaged condition.Both stiffeners were received loose within the returned package.During the investigation, one of the two stiffeners exhibited shaft damage and would not advance fully advance into the catheter.The second stiffener would not fully advance into the catheter, with no visual damage to the stiffener.The catheter i.D.And the o.D.Of the stiffeners were confirmed to be within specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot discovered one nonconformance confirmed to be relevant to the reported difficulty.These devices were scrapped prior to further processing.There were no recorded nonconformances for all of the related subassembly lots.To date, a further search of the database records revealed one additional complaint associated with this lot regarding a similar failure.The complaint was also reported from the same facility as this complaint.Cook also reviewed product labeling.The attached ifu, t_nucl_rev5, states under the heading "precautions": "when inserting a stiffening cannula into the catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of the suture." the information provided upon review of the dmr, dhr, ifu, and device evaluation suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, inspection of the returned device, and the results of the investigation, cook has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE COPE NEPHROURETEROSTOMY SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key19005811
Report Number1820334-2024-00442
Device Sequence Number18496893
Product Code FAD
UDI-Device Identifier00827002481701
UDI-Public(01)00827002481701(17)260613(10)15485953
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/13/2026
Device Model NumberN/A
Device Catalogue NumberULT8.5-8.5-20-NUCL-A-RH
Device Lot Number15485953
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by Manufacturer07/05/2024
Initial Report FDA Received Date03/29/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-