| Model Number |
3143 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Inadequate Pain Relief (2388)
|
| Date of Event |
02/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event estimated.During processing of this incident, attempts were made to obtain complete device information.Further information was requested but not received.During processing of this incident, attempts were made to obtain complete patient information.Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
It was reported that during an elective ipg replacement that took place on (b)(6) 2024.The system diagnostics recorded high impedances on the cervical lead and was not providing adequate therapy.The cervical lead is not currently connected to the new ipg, and surgical intervention may take place at a future date to address the lead issue and connect it to the new ipg.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event estimated.
|
| |
|
Event or Problem Description
|
|
Additional information was received stating that surgical intervention took place where the lead was explanted and replaced to address the issue.The ipg was also replaced but met longevity.Effective stimulation was restored.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
|
| |
|