| Catalog Number |
CL-07645 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "when injecting the contrast agent, it clogged at the sheath sidearm and did not flow.Therefore, the sheath was removed and replaced with a new one to complete the procedure.No patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported "when injecting the contrast agent, it clogged at the sheath sidearm and did not flow.Therefore, the sheath was removed and replaced with a new one to complete the procedure.No patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|