| Model Number |
21-345 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, prior to use, when the device was taken out from the box, the appearance of the device was scratched and damage was observed, so it was not used.No patient was involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection of the device noted no abnormalities to the foam body.The valve was opened.The surfactant was gone.Functionally, the tactile inspection of the power switch confirmed that the switch was fully activated.The device was dismantled, the foam body was torn to expose internal components.The device batteries were drained.Brown stain on the inside of the foam body and near the seal was observed.It was reported that the instrument broke.The reported issue was confirmed.The most likely cause was traced to device design.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|