• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PROTECTA VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PROTECTA VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D334VRG
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Codes Fall (1848); Shock from Patient Lead(s) (3162)
Date of Event 03/09/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3: date is approximate.Month and year are confirmed valid.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) inappropriately delivered shocks and anti-tachycardia pacing for atrial fibrillation (af) with rapid ventricular response when the device misclassified episodes as ventricular tachycardia (vt) and fast ventricular tachycardia (fvt).It was noted that the patient fell.The icd remains in use. no further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19006975
Report Number3004209178-2024-08154
Device Sequence Number8446278
Product Code LWS
UDI-Device Identifier00613994677846
UDI-Public00613994677846
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/28/2016
Device Model NumberD334VRG
Device Catalogue NumberD334VRG
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/29/2024
Date Device Manufactured02/02/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694765 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age77 YR
Patient SexMale
Patient Weight122 KG
-
-