| Model Number |
N/A |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that the involved glidesheath slender device sheath value leaked.Blood loss was less than 250 cc.Leakage occurred three times during the same patient heart cath, and the leakage was described as minimal.They replaced each gss with another of the same lot and each time it leaked again.There was no harm to patient who is stable condition.The event occurred intra-operative.There is not a direct allegation that the reported device caused or contributed to patient injury and/or need for medical intervention.The patient was not injured during the event and medical/surgical intervention was not needed.
|
| |
|
Additional Manufacturer Narrative
|
|
A1: patient identifier: requested, not provided a2: age or date of birth: requested, not provided a3a: sex: requested, not provided a3b: gender: n/a a4: weight: requested, not provided a5: ethnicity: requested, not provided a6: race: requested, not provided d6a: implanted date: device was not implanted d6b: explanted date: device was not explanted e3: occupation: lab manager the actual device is not available for return.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being sent as follow-up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for valve leakage because the complaint sample was not returned for assessment.The exact root cause could not be determined.Historically, valve leakages have been caused by off-center insertion of the dilator.Review of device history record showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the design failure mode and effects analysis (dfmea).
|
| |
|
Search Alerts/Recalls
|