• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC LAMINAR; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC LAMINAR; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number OPOHF21L
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: note: this mdr report is being submitted as part of a remedial action to prevent serious injury.Section a2, a3, a4, a5: not applicable as there was no patient involvement section b3: date of event: unknown/not provided.Section d6a: if implanted, give date: not applicable as this is not an implantable device.Section d6b: if explanted, give date: not applicable as this is not an implantable device.Section h3-other (81): the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Section h9: the jjsv reference number is (b)(4).The number was omitted due to the complaint handling system character limitation.Johnson & johnson surgical vision (jjsv) has initiated a field action related to select lots of laminar® flow irrigation sleeve and test chamber ¿ 21 gauge (part number: opohf21l) due to a manufacturing issue which could lead to subpar performance during cataract surgery.While not all the irrigation sleeves in the recalled lots are affected, this nonconformance could lead to insufficient flow of balanced saline solution (bss), which may result in an unstable anterior chamber.This defect could result in a capsule tear, iris trauma, or corneal edema, or in rare instances a corneal incision burn.Complaints will continue to be monitored and investigated.Investigation is ongoing and corrective action will be implemented as appropriate.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that account received a box of opohf21l with only 1 irrigation hole.There is no patient adverse event to report.There was no medical or surgical intervention required.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMINAR
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key19009674
Report Number3012236936-2024-00792
Device Sequence Number19803491
Product Code HQC
UDI-Device Identifier05050474573376
UDI-Public(01)05050474573376(17)260717(10)60479361
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K121721
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Expiration Date07/17/2026
Device Model NumberOPOHF21L
Device Catalogue NumberOPOHF21L
Device Lot Number60479361
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2024
Initial Report FDA Received Date04/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
SEE H10
Patient Sequence Number1
Patient SexUnknown
-
-