| Model Number |
EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the 78 y/o male patient had a revision due to rtsa was unstable and anteriorly dislocated within 1 week post op.Patient was revised tray/liner/torque screw to hat, a new liner, new torque screw.There was no breakage of the device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Images and x-rays received.The devices are not available for evaluation due to hospital would not decontaminate.No additional patient/medical information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of anterior total shoulder dislocation approximately one week post-op, which led to the damage on the explanted humeral liner.The reported dislocation could be confirmed from the provided radiographs.
|
| |
|