| Catalog Number |
5-12614 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "the doctor found cuff leakage when doing testing before using on patient." no patient invovlement occured.Threfore, no patient harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The reported complaint of "leak - device leak - cuff" was confirmed based upon the sample received.The returned et tube was found to have two parallel lines cut into the balloon cuff.There were multiple holes along the lines which would prevent the cuff from staying inflated.A device history record review was performed with no evidence to suggest a manufacturing related cause.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.Teleflex will continue to monitor and trend on complaints of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported "the doctor found cuff leakage when doing testing before using on patient." no patient invovlement occured.Threfore, no patient harm or injury.
|
| |
|