| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat# 650-1057 lot# 142150 cer bioloxd option hd36mm.Cat# 650-1066 lot# 571430 cer opt type 1 tpr sleve 0mm.Cat# unk lot# unk 5 cerclage wires.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a hip revision due to a right broken stem implant.There were no surgical complications.Though requested, no additional information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated.H6: component code: mechanical (g04) - stem.One stem was returned and evaluated.Upon visual inspection, it was noted that the device had fractured at the female taper bottom.There was visible gouging and scratches to the profile of the device, along with scuffing to the face and male taper shaft.The returned hip stem was submitted for fracture analysis.The fracture surface showed smearing and biological debris that obscured the artifacts.However, high magnification imaging revealed striations in multiple areas suggesting that the hip stem fractured due to fatigue.A review of the device history records identified no deviations or anomalies during manufacturing.The reported products were reviewed for compatibility with no issues noted.Medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues: right total hip arthroplasty with long stem from a component which appeared to be fractured at the junction of the femoral neck and proximal femoral stem.A definitive root cause cannot be determined.The event was confirmed via x-rays and returned product if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|