Model Number 101044164
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 03/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service conducted troubleshooting with the customer, including verifying breast shield fit.The customer was referred to a lactation consultant to verify breast shield size is correct and to get advice on a breast shield insert.The customer was sent new collection cups.The customer was contacted by a complaint handler to get additional information, with no response as of the date of this report.Based on the results of our internal investigation (reference number (b)(4)), it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 33% for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However, in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On 3/26/24, the customer alleged to medela llc that the collection cup for her freestyle handsfree pump was not closing properly and is leaking human breast milk.The customer also reported that she had developed mastitis and was prescribed antibiotics by her doctor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE HANDS-FREE EN ES FR - US ADAPTER
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA AG
lattichstrasse 4b
baar zug 6341
SZ  6341
Manufacturer (Section G)
MEDELA AG
lattichstrasse 4b
baar zug 6341
SZ   6341
Manufacturer Contact
tom cuzman
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key19011137
Report Number3002807523-2024-00005
Device Sequence Number20873582
Product Code HGX
UDI-Device Identifier00020451441641
UDI-Public020451441641
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K191653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101044164
Device Catalogue Number101044164
Device Lot Number0000600030
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/26/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/26/2024
Initial Report FDA Received Date04/01/2024
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-