MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 73022L
Medical Device Problem Code Material Invagination (1336)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley and syringe.Visual inspection of the sample noted using the returned syringe the catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1 percentage aqueous methylene blue per 100ml distilled water) and observed small ridge on the catheter balloon.This does not meet specification which sates "balloon must not cuff after deflation".The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be ¿balloon position length not established¿.However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: ¿bard® foley catheter warning: do not use ointments or lubricants having a petrolatum base.They will damage the catheter.Visually inspect the product for any imperfections or surface deterioration prior to use.- use luer tip syringe to inflate with stated ml of sterile water.Or - for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Contains or presence of natural rubber latex.Caution: this product contains natural rubber latex which may cause allergic reactions.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.¿ correction: d.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that while deflating the foley catheter balloon, a ring or ridge was created which caused pain to the patient during removal.No medical intervention was reported.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that while deflating the foley catheter balloon, a ring or ridge was created which caused pain to the patient during removal.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19011268
Report Number1018233-2024-01729
Device Sequence Number3879626
Product Code EZL
UDI-Device Identifier00801741034329
UDI-Public(01)00801741034329
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K070558
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number73022L
Device Lot NumberNGGY5148
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2024
Supplement Date Received by Manufacturer07/08/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-