Catalog Number 07K65-34
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Section b5 descibe event or problem was updated with: additional data for two more patients provided (b)(6) 2024: 1st sample = initial result = 2.37 repeat 1.13 ng/dl.2nd sample = initial result = 2.03 repeat 1.21 ng/dl.
 
Event or Problem Description
The customer observed falsely elevated architect free t4 results for four patients on architect i2000sr, serial numbers (b)(6).The results were lower when the samples were repeated.The following data was provided: (b)(6): (b)(6) 2024.Sample 1: initial test: >5 ng/dl.Retest: 1.22 ng/dl.(b)(6) 2024.Initial test: >5 ng/dl.Retest: 1.28 ng/dl.Another device: (b)(6): sample 3: initial test: 2.79 ng/dl.Retest: 1.19 ng/dl.Sample 4: initial test: 2.05 ng/dl.Retest: 1.31 ng/dl.Additional data for two more patients provided (b)(6) 2024: 1st sample = initial result = 2.37 repeat 1.13 ng/dl.2nd sample = initial result = 2.03 repeat 1.21 ng/dl no impact to patient management was reported.
 
Additional Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, device history record review, and labeling review.Additionally, in-house testing was performed.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue.A search for similar complaints did identify an increase in complaint activity, however, no related trends were identified regarding commonalities for lot number 56001ud00 and issue for the product.Additionally, in-house performance testing was completed which indicates the product is performing as expected.The device history record review did not identify any nonconformances, potential nonconformances or deviations associated with lot number 56001ud00 and the complaint issue.Labeling was reviewed and adequately addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the architect free t4 for lot 56001ud00 was identified.Section d4 catalog number and primary udi updated to (b)(4) with size code change from 24apr2024.
 
Event or Problem Description
The customer observed falsely elevated architect free t4 results for four patients on architect i2000sr, serial numbers (b)(6).The results were lower when the samples were repeated.The following data was provided: (b)(6): (b)(6) 2024 sample 1: initial test: >5 ng/dl retest: 1.22 ng/dl.(b)(6) 2024 sample 2: initial test: >5 ng/dl retest: 1.28 ng/dl.Another device: (b)(6): sample 3: initial test: 2.79 ng/dl retest: 1.19 ng/dl.Sample 4: initial test: 2.05 ng/dl retest: 1.31 ng/dl.Additional data for two more patients provided (b)(6) 2024: 1st sample = initial result = 2.37 repeat 1.13 ng/dl 2nd sample = initial result = 2.03 repeat 1.21 ng/dl no impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer observed falsely elevated architect free t4 results for four patients on architect i2000sr, serial numbers (b)(6).The results were lower when the samples were repeated.The following data was provided: (b)(6): (b)(6) 2024 sample 1: initial test: >5 ng/dl retest: 1.22 ng/dl (b)(6) 2024 sample 2: initial test: >5 ng/dl retest: 1.28 ng/dl another device: (b)(6): sample 3: initial test: 2.79 ng/dl retest: 1.19 ng/dl sample 4: initial test: 2.05 ng/dl retest: 1.31 ng/dl no impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT FREE T4 REAGENT KIT
Common Device Name
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key19011422
Report Number3005094123-2024-00149
Device Sequence Number15448885
Product Code CEC
UDI-Device Identifier00380740119805
UDI-Public(01)00380740119805(17)240926(10)56001UD00
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173122
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2024
Device Catalogue Number07K65-34
Device Lot Number56001UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/13/2024
Supplement Date Received by Manufacturer04/02/2024
04/24/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date04/17/2024
05/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2K PROC MOD, 03M74-01, (B)(6).; ARC I2K PROC MOD, 03M74-01, (B)(6).; ARC I2K PROC MOD, 03M74-01, (B)(6).
Patient SexUnknown
-
-