• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD SUZHOU (MDS) SAF-T-INTIMA W/Y ADAPTER YEL 24GA X .75I; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD SUZHOU (MDS) SAF-T-INTIMA W/Y ADAPTER YEL 24GA X .75I; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383313
Medical Device Problem Codes Retraction Problem (1536); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Dhr review: the complaint lot# is 2287484, sku is 383313, assembly in suzhou plant on 2022.Oct.29, lot quantity is (b)(4) ea.Review the in process test record and outgoing test report, all test results meet the product specifications, no abnormal found.Review the product assembly record, no non-conformities, deviations or rework activities for this lot.Returned sample analysis: no samples returned, there are two pictures attached, it shows a product within unit package, show product manufacture information only but not the defect sample retain sample analysis: sampling 2ea from the retain sample of the same lot to check product function, product safety shield activation function is good.Refer to the attachment for retain sample test report.Possible cause analysis: based on the reported information, needle is exposure to air.Possible reasons for this type of failure may including: 1.The assembly status between outer shield and rubber is not good, the rubber is not pressed to the end during assembly 2.Product safety shield is pulled during manual assembly flow or manual packaging.Thess risk will cause the outer shield drops off before the needle retracted completely.Current manufacture already has control procedures as below to defect and prevent this kind of defect: 1.100% inspection for the gap between outer and rubber is performed at the last station of assembly 2.Both in-process and outgoing sampling check will test the separation force between rubber and outer shield there is no defect sample returned, also no defect sample picture provided, without this, we cannot identify the actual defect feature, cannot determine whether it¿s a poor assembly issue or raw material issue, so the root cause of this case is not clear.Based on the ifu description, hold the puller and keep the component adown can activate needle safety function successfully.
 
Event or Problem Description
No additional information provided.
 
Event or Problem Description
"product fails with safety device, because at the end of the punch, when pulling the needle to activate the product's safety system, the water does not return completely, and when clamping to connect the serum, the needle is exposed, exposing the employee the risk of accident with drilling and sharp, a situation has already occurred in the sector.Product had never caused a problem before batch number 2287484, product purchased from industry itself, invoice (b)(4).Uninformed.
 
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAF-T-INTIMA W/Y ADAPTER YEL 24GA X .75I
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD SUZHOU (MDS)
#5 baiyu road
suzhou industrial park
jiangsu 21502 1
CH  215021
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou 21502 1
CH   215021
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19011512
Report Number3002601200-2024-00105
Device Sequence Number19782450
Product Code FOZ
UDI-Device Identifier00382903833139
UDI-Public(01)00382903833139
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K013800
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Catalogue Number383313
Device Lot Number2287484
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer04/11/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-