|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The cerelink monitor ((b)(6)) was returned for evaluation.Device history record (dhr) - the batch records were reviewed for any anomalies or report related to the finished device, and none were identified.Failure analysis - cerelink monitor requires rework to bring the unit to the current design state.Requires analog board replacement to correct the error code 1000.The rework performed includes the following: updating the digital board to correct the lcd on-state when the monitor is powered off and updating the lcd touch screen to the latest revision.Replaced the lithium coin battery.Replaced lithium monitor battery.Final functional testing and electrical safety testing performed.Testing completed and the monitor has passed all functional and electrical safety tests.Root cause analysis - the returned cerelink monitor required component upgrade to bring the unit to the current design state.The analog board was replaced to correct error code 1000.Final functional and electrical safety testing completed.
|