ZIMMER BIOMET, INC. ECHO BI-METRIC POR FMRL NC 11X135MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994)
Date of Event 03/07/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial hip procedure.Five days post op, the patient developed pain and was unable to bear weight.The patient had no trauma or falls post initial surgery.A secondary x-rays showed that the stem had subsided.Subsequently, the patient underwent a revision procedure approximately 2 weeks post implantation.During the secondary procedure, a fracture was found, cabled and a bigger stem was reamed, broached, and inserted.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: cat# 163670 lot# j7252771 32mm mod head cocr +3mm neck.Cat# 010000664 lot# 7586906 g7 pps ltd acet shell 54f.Cat# 010000928 lot# 7273204 g7 hi-wall e1 liner 32mm f.G2: foreign: new zealand.The customer has indicated that the product will not be returned to zimmer biomet for investigation as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04)- stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.The surgeon is unsure of when the bone fracture occurred, during the initial placement or sometime before the revision.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO BI-METRIC POR FMRL NC 11X135MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19012190
Report Number0001825034-2024-00867
Device Sequence Number7981562
Product Code KWA
UDI-Device Identifier00887868260421
UDI-Public(01)00887868260421(17)321212(10)387520
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K070274
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number192011
Device Lot Number387520
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer06/26/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date06/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-