MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL FREEFORM MINI 2MM X 6CM; CEREPAK FREEFORM MINI DETACHABLE COIL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL FREEFORM MINI 2MM X 6CM; CEREPAK FREEFORM MINI DETACHABLE COIL SYSTEM Back to Search Results
Catalog Number MCR092060
Medical Device Problem Codes Fracture (1260); Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the field, during a coil embolization, a freeform mini 2mm x 6cm coil (mcr092060, 31162594) was placed into a left middle cerebral artery (mca) bifurcation aneurysm.This was the fifth coil to be placed.The physician attempted to detach using manual break.Coil did not detach ¿ verified under fluoro.Proximal end of pusher tube detached from pull wire and slid off.The physician attempted to wrap pull wire around his finger and pull.Coil still would not detach ¿ verified under fluoro.Pull wire broke after pulling several times without successful detachment.The physician attempted to pull remaining portion of pull wire using hemostat.Coil still would not detach as verified under fluoro.Coil was removed from aneurysm and microcatheter without disturbance to coil mass.No injury to patient.Additional event information was received on 21-mar-2024 indicating that a stryker sl-10 microcatheter (mc) was used.There was no resistance during advancement of the device through the microcatheter or positioning difficulty in the aneurysm.There were no attempts using the detachment handle.There was no damage to the embolic coil or any device component.The vessel was moderately tortuous.The procedure was delayed by 5 minutes due to the multiple attempts to detach.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient race and ethnicity was not reported.Section d2b ¿ procode: krd/hcg.The device was discarded, therefore, no further investigate can be performed.A manufacturing record evaluation was performed for the finished device 31162594 number, and no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.The instructions for use (ifu) contain the following recommendations: 1.Remove the detacher from the protective packaging and place it within the sterile field.The detacher is packaged separately for single patient use only.2.Confirm again under fluoroscopy that the coil alignment marker of the delivery tube creates a ¿t¿ with the proximal marker of the microcatheter.3.Verify that the rhv is firmly locked around the delivery tube to ensure that the coil does not move during connection process.Ensure the delivery tube is straight between the rhv and the detacher.4.Hold the delivery tube approximately 3 inches from proximal end and bring the detacher over the delivery tube until a firm stop is felt.5.Maintain the delivery tube position within the detacher and use a thumb to pull back the slider on the detacher.6.Continue to pull on the slider until end of travel or a click is heard.Gently release the slider until it returns to starting position or a second click is heard.Remove detacher by slowly withdrawing it from the delivery tube.Note: detacher must remain in sterile field until the procedure is completed.7.Fluoroscopically verify that the coil is detached by gently pulling back the delivery tube approximately 1cm.If the coil has detached, remove the delivery tube from the microcatheter.8.If the coil has not detached, repeat steps 3-7.If 2 unsuccessful detachments occur, discard the detacher and proceed to the manual break section.9.Repeat steps 1-8 until the procedure is completed.Discard the detacher.Caution: always perform fluoroscopic verification of detachment prior to withdrawing the delivery system fully.Failure to do so may lead to an embolic complication.10.After detaching the coil, remove the delivery tube from the microcatheter and discard.11.Repeat the above sequence for all additional coils until the procedure is complete.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEFORM MINI 2MM X 6CM
Common Device Name
CEREPAK FREEFORM MINI DETACHABLE COIL SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS, INC.
calle circuito int norte 1820
parque industrial salvacar
juarez 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine 92618
3035526892
MDR Report Key19012333
Report Number3008114965-2024-00355
Device Sequence Number14458175
Product Code KRD
UDI-Device Identifier10886704083016
UDI-Public10886704083016
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K220040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue NumberMCR092060
Device Lot Number31162594
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2024
Initial Report FDA Received Date04/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STRYKER SL-10 MICROCATHETER.
Patient Age61 YR
Patient SexMale
Patient Weight83 KG
-
-