• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 4 MM, 12°, LONG; RIGID SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 4 MM, 12°, LONG; RIGID SCOPE Back to Search Results
Model Number WA20021A
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a therapeutic transurethral resection of prostate procedure, the telescope on its first use did not fail during the procedure, but it gave a view of the upper edge of the shaft and not coagulation loop.The patient was on general anesthesia and procedure was delayed (unspecified time) and could not be completed.The procedure was postponed and later cancelled to be scheduled for another day.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was confirmed that there was no patient injury due the reported event.However, the facility reported that consequently the patient will require a second round of anesthesia for the subsequent procedure.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the reported distorted image issue was not confirmed.However, the objective lens was displaced.Consequently, this caused an unclear image.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.Based on the results of the investigation, the reported distorted image was not confirmed.However, the displaced lens (which caused the unclear image) was likely due to excessive force by the user.This supplemental report includes a correction to b3, b5, and g2 to provide information that was inadvertently not included in the initial medwatch.An update has been made to d9 and h3.Also, information has been added to h4.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 4 MM, 12°, LONG
Common Device Name
RIGID SCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19014342
Report Number9610773-2024-00918
Device Sequence Number12760850
Product Code GCP
UDI-Device Identifier04042761051422
UDI-Public04042761051422
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K897003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA20021A
Device Lot Number800962
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer04/17/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-