• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKNSON AND CO. - DUN LAOGHAIRE CO, IRELAND PEN NDL 29GA 12.7MM 100 BX 1200 USA; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKNSON AND CO. - DUN LAOGHAIRE CO, IRELAND PEN NDL 29GA 12.7MM 100 BX 1200 USA; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 328203
Medical Device Problem Code Misconnection (1399)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported, she is not able to attach the non patient end to her insulin pen prior to injection.Stated, she is noticing an "extra piece of plastic" in the hub that is preventing the non patient end from attaching.Lot: 1258948; catalog: 328203; date of event: unknown; samples: yes cl.
 
Additional Manufacturer Narrative
Additional information was added to: b4, g6, h2, h3, h11.Correction to: d3 (country type and country), h6(component code, investigation findings, and investigation conclusions) investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NDL 29GA 12.7MM 100 BX 1200 USA
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
BECTON DICKNSON AND CO. - DUN LAOGHAIRE CO, IRELAND
EI 
Manufacturer (Section G)
BECTON DICKINSON AND CO.
1 becton drive
franklin lakes, NJ 07417
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key19015288
Report Number9616656-2024-05205
Device Sequence Number3880276
Product Code FMI
UDI-Device Identifier00382903282036
UDI-Public00382903282036
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K213478
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2026
Device Model Number328203
Device Lot Number1258948
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date06/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-