• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 18GA X 10CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 18GA X 10CM Back to Search Results
Model Number 360-1080-02
Medical Device Problem Code Failure to Obtain Sample (2533)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.One opened sample was returned from the customer for review.A visual inspection was performed on the returned product, and it was found that the pincer was flattened, inhibiting it from properly obtaining a sample.The complaint was confirmed.There are stringent computer and photographic inspections that ensure the pincer is not deformed during the manufacturing process and has the proper radius.An internal investigation has been opened to investigate the issue in detail and it is under effectiveness phase.A review of the complaints database was conducted, and over the previous six months, multiple similar complaints were found related to this part number 360-1080-02.There were multiple similar complaints found related to this lot number.
 
Event or Problem Description
An ultrasound manager reported that during a kidney biopsy, 4 passes were performed with no specimen received.There was no patient injury.
 
Additional Manufacturer Narrative
Udi related data quality updates only.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 18GA X 10CM
Common Device Name
BIOPINCE ULTRA
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
kelly mayer
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key19015333
Report Number0001625425-2024-00973
Device Sequence Number19803541
Product Code KNW
UDI-Device Identifier00886333224005
UDI-Public(01)00886333224005(17)260613(10)11482911
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K904987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/13/2026
Device Model Number360-1080-02
Device Lot Number11482911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer09/04/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-