MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0720
Medical Device Problem Code Labelling, Instructions for Use or Training Problem (1318)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.The subject device was manufactured on september 2022 based on the provided lot information.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
Based on the device evaluation, it is likely this single-use sphincterotome was used multiple times.There were no reports of patient involvement.
 
Event or Problem Description
Additional information from the customer stated that the procedure was therapeutic with no delay and was completed without patient injury nor harm.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the event (reused disposable products) was confirmed, and the device did not meet the standard specifications.It was determined that the suggested event was caused by device handling.The device was inspected and was found to have been used, which was suggestive of black colored wire that was intact.At bow position (pulled), the wire was straight and in correct position.While attempting de-bowing the wire came out of the plane of the sheath, which was determined to be due to excessive push.The instruction manual identifies the following related verbiage which could have prevented the phenomenon: (drawing no.Gk6226 revision no.13): ·do not attempt to sterilize the instrument.Doing so could pose an infection control risk, cause tissue irritation, equipment damage, or malfunction.·the instrument is a single-use, disposable item.Do not reuse or attempt to sterilize it.Reusing the instrument could pose an infection control risk, cause tissue irritation or malfunction.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19015638
Report Number9614641-2024-00811
Device Sequence Number15494479
Product Code KNS
UDI-Device Identifier04953170184017
UDI-Public04953170184017
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberKD-V411M-0720
Device Lot Number29K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer04/25/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-