| Catalog Number |
29MECJ-502 |
| Medical Device Problem Code |
Perivalvular Leak (1457)
|
| Health Effect - Clinical Code |
Aortic Valve Insufficiency/ Regurgitation (4450)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was noted that on (b)(6) 2024, a 29mm sjm masters series valve expanded cuff was chosen for a procedure.The valve was implanted.One minute after the implantation, a moderate valvular leak was noted.The valve was explanted and mitral valve replacement (mvr) was done with a new 29mm sjm masters series valve expanded cuff intra-procedurally, while the patient was still on bypass.The patient remained hemodynamically stable throughout the procedure.There was no delay in the procedure.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant due to perivalvular leak was reported.The investigation found that both leaflets were intact and opened and closed completely without difficulty.No material was seen within the pivot recesses or on the surfaces of the mechanical leaflets.The sewing cuff was intact.No inflammation.Hydrodynamic testing to evaluate valve function.Data and video of valve function, along with open & closed images, were obtained to evaluate hydrodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received the cause of the reported incident could not conclusively be determined.
|
| |
|
Event or Problem Description
|
|
It was noted that on (b)(6) 2024, a 29mm sjm masters series valve expanded cuff was chosen for a procedure.The valve was implanted.One minute after the implantation, a moderate valvular leak was noted around the valve.The valve was explanted and mitral valve replacement (mvr) was done with a new 29mm sjm masters series valve expanded cuff intra-procedurally, while the patient was still on bypass.The patient remained hemodynamically stable throughout the procedure.There was no delay in the procedure.The patient was reported stable.
|
| |
|
Search Alerts/Recalls
|