MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN SCW POS WG TRL10M SZ5-6LG; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN SCW POS WG TRL10M SZ5-6LG; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71431806
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a revision knee procedure performed on (b)(6) 2024 to remove a tka system from another company, the surgeon required the use of augments for the femur and tibia.While assembling the femoral posterior augment, the doctor realized there was difficulty in securing the screw.It was then noticed that part of the screw of a lgn scw pos wedge trial10m sz5-6lg had come away and was left in the lgn ox constrain femoral trial 6 lt.The doctor was able to complete the use of the trial while the screw part was still attached.The procedure was successfully completed after a non-significant delay.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
H10: internal complaint reference case (b)(4).
 
Additional Manufacturer Narrative
Additional information: h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: the wedge trial was not returned for evaluation; therefore, a device analysis could not be performed.However, the photographs were reviewed and revealed that the wedge's bottom screw detached from the device and was left inside hole of the femoral trial.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Corrected data: h6 (health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN SCW POS WG TRL10M SZ5-6LG
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19016766
Report Number1020279-2024-00642
Device Sequence Number7981196
Product Code HRY
UDI-Device Identifier03596010547385
UDI-Public03596010547385
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71431806
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer04/22/2024
Initial Report FDA Received Date04/01/2024
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LGN OX CNSTR FEM TRL6 LT(PN:71431166,LN:07FM15574).
Patient SexUnknown
-
-