| Catalog Number |
200009902 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Inadequate Osseointegration (2646); Limb Fracture (4518)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery due to tibial component loosening and polyethylene debris.The left ankle was aspirated today through a anteromedial approach and fluid was sent for total cell count, gram stain, culture and sensitivity.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, based on available ct scans and hcp assessment.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans, healthcare professionals opined that- ¿tibial (construct) ¿ the tibial construct shows clear radiolucence and large cysts around the stem.The cysts partially destroyed the cortex laterally and led to a periprosthetic erosion (fracture like-lesion).Loosening, but no clear migration can be confirmed with this finding.Pe ¿ the pe is not separated from the tibial tray.However, there is nothing visible in the ct, but the report suspects, that loosening may be the cause of pe debris.Talar (construct) ¿ there is some, rather insignificant radiolucence.Full tar (construct) ¿ the failure might have been partly caused by the device through debris.The cause of the debris, though, is not clear.There are also further investigations ongoing to exclude signs of infection.An infection would be more patient related.If there is significant debris of the pe, this should be confirmed when the implant is observed and with additional histological material from the bone/joint.¿ based on the investigation, the root cause was attributed to a patient related issue.The event was caused by presence of cysts around the tibial construct which resulted in loosening of tibial construct.Furthermore, the cysts partially destroyed the cortex laterally and led to a periprosthetic erosion (fracture like lesion).
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery due to tibial component loosening and polyethylene debris.The left ankle was aspirated today through a anteromedial approach and fluid was sent for total cell count, gram stain, culture and sensitivity.
|
| |
|