• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TURNCARE, INC. GUARDIAN 2 SYSTEM; ALTERNATING PRESSURE SUPPORT SURFACE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TURNCARE, INC. GUARDIAN 2 SYSTEM; ALTERNATING PRESSURE SUPPORT SURFACE Back to Search Results
Model Number GS2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pressure Sore/Ulcer (2326)
Date of Event 04/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted based on a retrospective review (reference rca # lvhn ipcu).Investigation was conducted on the reported information and from the data collected by the guardian device available at the time.
 
Event or Problem Description
Patient was admitted on (b)(6) 2023.Patient's stage 2 on natal cleft progressed to unstageable on (b)(6) 2023.Evaluation of guardian data was compiled for dates prior to hapi discovery.A total gap in therapy of 143.5 hours was reviewed, which included: 12.5 hours of disconnected enhancer, 92 hour delay in therapy when transferring between units, 34.5 hours of no patient detected/mispositioned enhancer, and 4.5 hours of the pause feature running for the full 30 minutes.There were also 14.5 total days utilizing the device with an incorrect weight selection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUARDIAN 2 SYSTEM
Common Device Name
ALTERNATING PRESSURE SUPPORT SURFACE
Manufacturer (Section D)
TURNCARE, INC.
230 west parkway unit 6
pompton plains NJ 07444
Manufacturer (Section G)
TURNCARE, INC.
230 west parkway unit 6
pompton plains NJ 07444
Manufacturer Contact
richard kalita
230 west parkway unit 6
pompton plains, NJ 07444
MDR Report Key19018688
Report Number3013682457-2024-00003
Device Sequence Number15496884
Product Code FNM
UDI-Device Identifier00860001236453
UDI-Public(01)00860001236453
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberGS2
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2023
Initial Report FDA Received Date04/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
Patient Weight55 KG
-
-